FREQUENCY
Frequency Range 0.01 to 21 GHz in six selectable ranges
Frequency Spans
Fullband displays entire spectrum of selected band
Per Division 10 kHz to 200 MHz/div in a 1, 2, 5 sequence
Zero Span analyzer functions as a manually tuned receiver
Frequency Accuracy
Tuning Accuracy 0.01 to 3 GHz: < ± (1 MHz + 0.3% of center frequency); 3 to 21 GHz: < ± (5 MHz + 0.2% of center frequency)
Frequency Span Accuracy < ± 5% of displayed frequency separation
Spectral Resolution
Resolution Bandwidths eight selectable resolution (3-dB) bandwidths from 1 kHz to 3 MHz in a 1, 3 sequence. Bandwidth and frequency span are independently variable or may be coupled for optimum display when control markers are aligned.
Resolution Bandwidth Accuracy 3-dB points are < ± 15% (except for 3 MHz bandwidth: ± 30% )
Selectivity (60-dB/3-dB bandwidth ratio) < 15:1
Spectral Stability (fundamental mixing, bands 0.01-3 GHz and 6-9 GHz)
Residual FM < 1 kHz p-p in 0.1 second
Noise Sidebands 70 dB down, 30 kHz from center of CW signal with 1 kHz resolution bandwidth and full video filtering
AMPLITUDE
Amplitude Range -111 to +30 dBm
Maximum Input (damage) Levels
Total Power +20 dBm (100 mΩ, 2.2 Vrms) with 0 dB input attenuation; +30 dBm (1 watt, 7.1 Vrms) with 10 dB input attenuation
dc or ac (< 100 Hz) ± 7.1 V
Peak Pulse Power + 50 dBm (100 Ω, < 10 ms pulse width, 0.01% duty cycle) with 30 dB input attenuation
Gain Compression < 0.5 dB for -10 dBm signal, 0 dB input attenuation
Average Noise Level Frequency Range (GHz) Avg. Noise Level (dBm/1 kHz) Frequency Response (± dB max.) Amplitude Accuracy (± dB max.)
0.01-3
6.0-9
3.0-9
9.0-15
6.0-15
12.1-18
18.0-21
-111
-108
-103
-98
-93
-92
-90
1.0
1.0
1.5
1.8
2.1
2.3
3.0
2.3
2.3
2.8
3.1
3.4
3.6
4.3
Alternate IF regular IF at 3.0075 GHz; alternate IF available at 2.9925 GHz for all frequency bands
Calibrated Display Range
Log 70 dB with 10 dB/div scale; 8 dB with 1 dB/div scale
Linear 8 divisions with linear (LIN) amplitude scale
Amplitude Accuracy
Calibrator -10 dBm ± 0.3 dB (into 50 Ω), 35 MHz ± 400 kHz
Reference Level 10 dB steps and a 12 dB vernier for calibrated adjustment from -112 dBm to +60 dBm
Step Accuracy (with 0 dB input attenuation) -10 to -80 dBm: ± 0.5 dB, -10 to -100 dBm: ± 0.1 dB
Vernier Accuracy ± 0.5 dB
Frequency Response see table above; includes input attenuator mixer flatness, and mixing mode gain variation (band to band), with 0 1or 10 dB input attenuation
Input Attenuator 0 to 70 dB, selectable in 10 dB steps
Step Accuracy < ± 1.0 dB per 10 dB step (0 to 60 dB, 0.01 to 18 GHz)
Maximum Cumulative Error < ± 2.4 dB (0 to 60 dB, 0.01 to 18 GHz)
Bandwidth Switching (Amplitude Variation)
3 MHz to 30 kHz < ± 0.5 dB
3 MHz to 1 kHz < ± 1.0 dB
Display Fidelity
Log Incremental Accuracy ± 0.1 dB/dB from Reference Level
Log Maximum Cumulative Error ± 1.5 dB over 70 dB range
Linear Accuracy ± 3% of Reference Level
Spurious Responses
Second Harmonic Distortion > 70 dB below a -40 dBm signal with 0 dB input attenuation
Third Order Intermodulation Distortion > 70 dB below two -30 dBm input signals (> 5 MHz) separated by 50 kHz and with 0 dB input attenuation
Residual Responses < -90 dBm with 0 dB input attenuation and no signal present at input (0.013 -3 GHz, 6-9 GHz)
SWEEP CHARACTERISTICS
Automatic sweeptime is automatically adjusted to maintain absolute amplitude calibration for any combination of frequency span, resolution bandwidth, and video filter bandwidth
Calibrated Sweep Times 2 msec to 10 sec/div in a 1, 2, 5 sequence (except 2 sec/div), ± 10% accuracy (± 20% for 5/10 sec/div)

to 10 sec/div in 1, 2, 5 sequence with ± 10% typical accuracy

Manual Sweep spectrum analyzer may be swept manually in either direction with front panel control
SIGNAL INPUT CHARACTERISTICS
Input Impedance 50 Ω  nominal; precision Type-N female connector
Input SWR typically < 2.0, 0 dB input attenuation < 1.3, 10 dB input attenuation
OUTPUT CHARACTERISTICS
Vertical Output, AUX A BNC output (50 Ω) detected video from 0 to 0.8 V for 8 division deflection on CRT display
Penlift/Blanking, AUX B BNC output provides 0 V pen down/unblanking signal at low impedance; 15 V penlift/blanking at 10 kΩ  impedance
21.4 MHz IF Output, AUX C BNC output (50 Ω) provides a signal which is proportional to the RF input. Level is about -10 dBm (into 50 Ω) with a signal displayed at the Reference Level. Output is controlled by settings of Resolution BW, Input Atten, and Reference Level
Horizontal Output, AUX D BNC output (5 kΩ) provides horizontal sweep from -5 V to +5 V for full 10 division CRT horizontal deflection
 
HP 8559A Product Page
 

Used Equipment Catalog

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved