Test Equipment Depot - 800.517.8431

View Shopping Cart | My Account

 

Free Ground Shipping for Orders over $100

Used Test Equipment
Completely refurbished,
aligned & calibrated to original
manufacturers' specifications
Used Equipment Catalog
<< HP 3580A Product Page

Hewlett Packard 3580A
5 Hz - 50 kHz Spectrum Analyzer Specifications

FREQUENCY CHARACTERISTICS
Range 5 Hz to 50 kHz
Digital Frequency Display
Resolution ± 1 Hz
Accuracy ± 3.5 Hz, 0 to 55° C
Typical Stability ± 10 Hz/hr after 1 hour; ± 5 Hz/°C
Bandwidths (accuracy ± 15%) 1 Hz (25°C ± 5C), 3 Hz, 10 Hz, 100 Hz, 300 Hz
Shape Factor 1 Hz (25°C ± 5C), 3 Hz, 10 Hz, 100 Hz: 10
300 Hz: 8
Out of range blank: if controls are set so portions of displayed signal lie below 0 Hz or above 50 kHz, the baseline is displayed.
AMPLITUDE CHARACTERISTICS
Overall Instrument Range Linear: 20 V -100 nV full scale
Log: +30 dBm or dB V;
        -150 dBm or dB V
Amplitude Accuracy Log Linear
Frequency response
     20 Hz - 20 kHz ± .3 dB ± 3%
     5 Hz - 50 kHz ± .5 dB ± 5%
Switching between bandwidths (25°C)
     3 Hz - 300 Hz ± .5 dB ± 5%
     1 Hz - 300 Hz ± 1 dB ± 10%
Amplitude display ± 2 dB ± 2%
Input attenuator ± .3 dB ± 3%
Amplitude reference level: (IF attenuator)
Most sensitive range ± 1 dB ± 10%
All other ranges ± 1 dB ± 3%
Dynamic Range 80 dB
IF feedthru input level > 10V, -60 dB; < 10V, -70 dB
Spurious responses > 80 dB below input reference level
Smoothing 3 positions, rolloff is a function of bandwidth
Overload indicator this LED indicator warns of possible input amplifier overloading. Without this indication it would be possible to introduce spurious responses without knowing it.
SWEEP CHARACTERISTICS
Scan Width 50 Hz to 50 kHz
Log Sweep 20 Hz to 43 kHz ± 20% after 3 sweeps
Sweep Times 0.1 s to 2000 s
Rep repetitive sweeps over the specified band
Reset resets to the beginning of the sweep-used to adjust start or center frequency
Manual in combination with the concentric knob, manual sweep fully duplicates the span of the electronic sweep
Adaptive Sweep when in adaptive sweep below the threshold level, scan speed is 20 to 25 times faster. Threshold is adjustable to cover 0-60% of screen. Signals greater than about 6 dB above threshold are detected and swept slowly.
Sweep Error Light this LED indicates a sweep that is too fast to capture full response. When the light is on, response can be > 5% lower than it should.
Zero Scan to look at the time varying signal at the center or start frequency within the bandwidth selected, the zero scan is used.
OUTPUT CHARACTERISTICS
Tracking Generator Output (also known as BFO or tracking oscillator output)
Range 0 to 1 V rms into 600 Ω
Frequency Response ± 3%, 5 Hz to 50 kHz
Impedance 600 Ω
Total Harmonic and Spurious Content 40 dB below 1 volt signal level
X-Y Recorder Analog Outputs
Vertical 0 to +5 V ± 2.5%
Horizontal 0 to +5 V ± 2.5%
Impedance 1 kΩ
Pen Lift contact closure to ground during sweep
Recommended Accessory HP 7090A Measurement Plotting System
Size 203.2 mm H X 285.8 mm W X 412.8 mm D (8" X 11 1/4 X 16 1/4")
Weight net, 12.25 kg (27 lb); HP 3580A Opt 001: net, 15.88 kg (35 lb)
Temperature Range 0°C to 55°C
Power 100V, 120V, 220V, or 240V +5% -10%; 48 to 440 Hz; 35 VA max
Opt 001 battery 5 hours from full charge. 14 hours to fully recharge. The internal battery is protected from deep discharge by an automatic turn off. Useful life of batteries is over 100 cycles.
 
3580A Product Page
 

Used Equipment Catalog

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved