500MHz Network/Spectrum Analyzer Specifications

Source
NETWORK MEASUREMENT
Frequency 10Hz to 500MHz, 1mHz resolution
Power -50dBm to +15dBm, 0.1dB resolution
Sweep Parameters Frequency, power and dc bias level
Sweep Types Linear, log. cw, program and partial
Output 2 outputs
DC bias level ±40V, 10mV resolution
Receiver
Frequency 10Hz to 500MHz
Input 4 inputs, 50 Ω  nominal
Resolution Bandwidth 3Hz to 300kHz, 1 or 3 step
Input Crosstalk -100dB
Magnitude Ratio
Dynamic Range > 100dB
Resolution 0.001dB
Dynamic Accuracy (23 ± 5°C), -30dBm R input ±0.05dB @ -70dBm to -30dBm T input
Phase
Range ±180°
Resolution 0.01°
Dynamic Accuracy (23 ± 5°C, -30dBm input) ±0.3° @ -70 to -30dBm T input
Delay
Range 10ps to 500s
Resolution 10ps @ 3.6 MHz aperture
Accuracy depends on phase accuracy
Error Comprehension
Mode Normalization, 1 port partial cal, 1 port full cal and port extension
SPECTRUM MEASUREMENT
Frequency
Measurement Range 10Hz to 500MHz
Resolution RBW: 3Hz to 300kHz, 1 or 3 step
Selectivity (60/3dB): 
4 for 3Hz to 30Hz
8.5 for 100Hz
Amplitude
Measurement Range -135 dBm to +20dBm
Accuracy ±1.0dB @ 50MHz
Linearity (23 ± 5°C) ±0.1dB @ -40 to 0dB; ±0.2dB @ -60 to -40dB
Frequency Response ±1.5dB
Dynamic Range (23 ± 5°C)
Second Harmonic Distortion -70dBc @ ³ 2MHz
T.O.I. Distortion ≤-80dBc @ ³ 2MHz
Residual Response -110dB @ ³ 100kHz
Average Noise Level typically -140dBm @ 10Hz RBW, ³ 2MHz
Sweep
Sweep Type Linear, log, cw, program and partial
Sweep Mode Continuous, single and manual
Sweep Time approximately 500 ms/span to 1800 s/span 
Input
Number of inputs 4 inputs
Impedance 50 Ω nominal
Damage level +30 dBm
Attenuator 0 to 50dB, 10dB step
DISPLAY AND ANALYSIS
Display 7.5 inch color CRT
Display Format Rectangulars, Table, Smith and Polar
Traces 4 traces max
Scale Type Linear, log
Autoscale
Phase display Expansion Display phase continuously more than ± 180 deg.
Video Filter Digital video filtering reduces random noise
Comment Entry Display a comment used alphabet, numeral and special characters (, , %, etc.)
Marker MKR ® Max (Min, Ref, Center, Start and Stop), Next Peak, Width and Delta reading mode
USER FUNCTIONS
User Math Puts the result in the form needed for your application by using built-in math operators, arithmetic functions and editing capability.
User Defined Function Provides one-key solution for a specific application without an external computer.  6 user functions can be created and soft-keys can be labeled as you like.
User Program (Auto Sequence Program) Allows to program the control or measurement, analysis, copy and other sequence without an external computer.
HARDCOPY
Dump Mode Copy the CRT display to a graphics printer
Plot Mode Copy the traces, graticule and annotations to an HP-GL compatible digital plotter
Print Mode Copy measurement data in tabular form to a printer
Hardcopy of traces, measurement data, results of analysis and annotations are produced by the 4195A and HP plotters or printers with LISTEN only capability.
STORAGE
Instrument state, trace data, table of Program Sweep and User Program can be independently saved or recalled from the built-in 3.5 inch floppy disk memory via SAVE/GET function.
Instrument state includes active control setting of measurement, active calibration data, active display format, active scale setting, User Math and User Define Function.
REMOTE PROGRAMMING
Interface Function SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1
Transfer Formats ASCII
32/64 bit IEEE 754 floating point format
HP-IB interface operates according to IEEE 488-1987 and IEC 625 standards and IEEE 628-1982 recommended practices.
GENERAL CHARACTERISTICS
Operating Conditions
Temperature 0°C to +45°C
Humidity 95% RH at 40°C
Non-Operating Conditions
Temperature -40°C to +70
Safety Based on IEC-348, UL-1244
Power 100, 120, 220V ±10%, 240V -10% +5%, 48Hz to 60Hz, 500VA (max)
Dimensions 425 (W) X 375 (H) X 620 (D) mm
Weight approximately 41kg
 

Used Equipment Catalog

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved