Hewlett Packard 3577A
5Hz - 200 MHz Network Analyzer Specifications

SOURCE
Frequency
Range 5 Hz - 200 MHz
Resolution 0.001 Hz
Amplitude ± 5 X 10 -8 /day, 0 to 55°C
Range +15 to -49 dBm (1.26 Vrms to 793m Vrms: 2dBV to -62 dBV) into a 50Ω  load
Resolution 0.1 dB
Accuracy ± 1 dB at +15 dBm and 100 kHz. Below +15 dBm, add the greater of ± 0.02 dB/dB or 0.2 dB
Flatness 1.5 dBp-p from 5 Hz to 200 MHz
Impedance 50Ω; > 20 dB return loss at all levels
RF Output Connector 50Ω  Type N female
Sweep Types Linear, alternate, cw and log frequency; log amplitude
Sweep Time 100 ms/span to 200 ms/span for frequency sweep; 1 ms/step to 16 s/step for amplitude sweep
Sweep Modes Continuous, single, manual
Trigger Modes Free run, immediate, line, external
RECEIVER
Full Scale Input Level -13 dBV from 10 kHz to 200 MHz with internal 20 dB attenuators ON (0dBm at 50Ω)
INPUT CHARACTERISTICS
Frequency Range 5 Hz - 200 MHz
Inputs 3 (A, B, R)
Input Impedance Selectable 50Ω  with > 25 dB return loss, or 1 MΩ  in parallel with approximately 30 pF
Input Connectors 50Ω  Type N female
Resolution Bandwidth Selectable 1 kHz, 100 Hz, 10 Hz, or 1 Hz
Sensitivity (Due to noise and internal crosstalk between source and receiver inputs) Resolution Bandwidth 30 kHz - 200 MHz (50Ω)
30 kHz - 20 MHz (1 MΩ)
Internal 20 dB Attenuator ON Internal 20 dB Attenuator OFF
1 Hz
10 Hz
100 Hz
1 kHz
-110 dBm
-110 dBm
-105 dBm
-95 dBm
-130 dBm
-130 dBm
-125 dBm
-115 dBm
Crosstalk > 100 dB isolation between inputs
Electrical Length/Reference Plane Extension Provides equivalent electrical line length, or delay at inputs A, B, and R.
Range: -3 X 108 m to +3 X 108 m or +1 to -1 s
Resolution 5 digits or 0.1 cm (3.3 Ps) whichever is greater
Accuracy ± 0.1 cm or ± 0.02% whichever is greater
MAGNITUDE CHARACTERISTICS
Range Full Scale Input to Sensitivity
Resolution Marker: 0.002 dB (log); 5 digits (linear)
Display: 0.01 dB/div to 20 dB/div (log absolute); 0.01 dB/div to 200 dB/div (log ratio); 0.1 NV/div to 10 V/div (linear absolute); 10-10/div to 1020/div (linear ratio)
Display Units dB, DBMS, div, V, and linear ratio
Accuracy (at 100 kHz, 25°C, and Full Scale Input)
Absolute (A, B, R) ± 0.2 dB
Ratio (A/R, B/R, A/B) ± 0.15 dB (50Ω); ± 0.2 dB (1 MΩ)
Dynamic Accuracy

Error

Input Level Relative to Full Scale Input

Resolution Bandwidth

1 kHz, 100 Hz, 10 Hz 1 Hz
± .04 dB
± .02 dB
± .05 dB
± .15 dB
± .75 dB
± .75 dB
± .04 dB
± .02 dB
± .05 dB
± .25 dB
± .75 dB
± 3.00 dB
0 dB to -10 dB
-10 dB to -50 dB
-50 dB to -60 dB
-60 dB to -80 dB
-80 dB to -90 dB
-90 dB to -100 dB
Frequency Response (when driven from a 50Ω  source and with a 50Ω receiver input impedance)
Absolute (A, B, R) 0.3 dBpp from 20 Hz to 20 MHz; 0.6 dBpp from 5 Hz to 200 MHz
Ratio (A/R, B/R, A/B) 0.3 dBpp from 20 Hz to 20 MHz; 0.4 dB from 5 Hz to 200 MHz
Reference Level Range: -207 DBMS to +33 DBMS (-220 div to +20 div) (Log absolute); -400 dB to +400 dB (log ratio); 0 V to 10 V (linear absolute); 0 to 1020 (linear ratio)
Stability Temperature: Typically <± 0.02 dB/°C
Time: Typically <± 0.05 dB/hour at 25°C
PHASE CHARACTERISTICS
Range ± 180 degrees
Resolution Marker: 0.005 deg (0.001 rad)
Display: 0.01 deg/div to 200 deg/div (0.00018 rad/div to 3.49 rad/div)
Accuracy (at 100 kHz, 25°C, and Full Scale Input) ± 2.0°
Dynamic Accuracy

Error

Input Level Relative to Full Scale Input
± .4 deg
± .2 deg
± .5 deg
± 1.5 deg
± 7.5 deg
0 dB to -10 dB
-10 dB to -50 dB
-50 dB to -60 dB
-60 dB to -80 dB
-80 dB to -100 dB
Reference Level Resolution 0.01°
Temperature Stability Typically <± 0.05 deg/°C
Time Stability Typically <± 0.05°/hour at 25°C
POLAR DISPLAY CHARACTERISTICS Range, Resolution, Display Units, Dynamic Accuracy, Frequency Response, Uncertainty, Crosstalk, Reference Level, and Stability specifications are the same as the corresponding magnitude and phase characteristics.
Full Scale Magnitude Range Absolute (A, B, R): 0.1 nV to 10 V
Ratio (A/R, B/R, A/B): 10-10 to 1020
REAL, IMAGINARY DISPLAY CHARACTERISTICS Range, Dynamic Accuracy, Frequency Response, Uncertainty, Crosstalk, Stability specifications are the same as the corresponding magnitude and phase characteristics.
DELAY CHARACTERISTICS
Range 1 ps to 20,00s
Resolution .01ns/div to 1000s/div
Normalized Accuracy Dynamic Phase Accuracy   +2nS
  360 X Aperture [Hz]
Aperture Range 0.5% to 16% of frequency span
Reference Level ± 103 S
GENERAL DISPLAY CHARACTERISTICS
Traces
No. Traces Two simultaneous traces may be present with a rectangular graticule. One trace with polar or Smith graticules.
Markers: Each trace has one main marker and an offset marker.  Markers indicate data at corresponding trace coordinates in the same units as used to set the Reference Level.   Markers can be used to modify certain display parameters.  Marker resolution is the same as horizontal display function.
Graticules
Rectangular Graticule 0% to 100% full scale deflection on 0.05% increments.  Logarithmic and Linear.
Polar/Smith Chart Graticule ± 500 deg in 0.001 deg increments
Noise Averaging
Type Exponentially weighted vector averaging on successive sweep data
Averaging Factor Selectable 1 (off), 4, 8, 16, 32, 64, 128, 256
Linear Phase Slope Compensation Provides linear phase slope offset of -72,000 deg/span to +72,000 deg/span
Calibration
Transmission Both traces can be normalized to measured data with full accuracy and resolution.
Reflection Corrects for directivity, frequency response and source match errors
PROGRAMMING CHARACTERISTICS
Capability Remote programming via the Hewlett-Packard Interface Bus (HP-IB). The HP 3677A/B S-Parameter Test Sets are programmable through the HP 3577A interface only.
Interface Functions SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP1, DC1, DT1, C0, E1
Output Data Transfer Time 401 data points (single parameter) can be transferred directly to an HP 200 series computer in Basic language as follows:
ASCII mode: Typically 1500 ms
Binary-floating point mode: Typically 160 ms
Graphics Capabilities 12 lines of text with 40 alphanumeric characters per line, and high resolution line vectors can be displayed through HP-IB commands
GENERAL CHARACTERISTICS
External Reference Frequency Input
Frequency 10 MHz/N. N is an integer from 1 to 100.
Level 0 dBm ± 10 dB, nominal
Impedance 50Ω, nominal
Connector BNC female, rear panel
Reference Frequency Output
Frequency 10 MHz
Level Typically 0 dBm
Impedance 50Ω, nominal
Connector BNC female, rear panel
External Trigger Triggers on negative TTL transition or contact closure to ground
Connector BNC female, rear panel
Plotter Control Directly compatible with HP-IB graphics plotters that use Hewlett-Packard Graphics Language (HP-GL) with listen only capability: HP 7470A, HP 7475A, HP 7550A, HP 7090A.
Save/Recall Front-panel setups can be stored in non-volatile memory locations 1 through 5. Last state is saved when power is removed.
Operating Conditions
Temperature 0 °C to +55 °C
Relative Humidity <95% at 40 °C
Altitude <4,572 m (15,000 ft)
Non-Operating Conditions
Temperature -40 °C to +75 °C
Altitude <15,240 m (50,000 ft)
Power 115V + 10%, -25% (47 Hz to 440 Hz), or 230 V + 10%, -15%
Weight 31 kg (67 lb) net; 41 kg (90 lb) shipping
Dimensions 222 mm H X 426 mm W X 578 mm D ( 8.75 in. X 16.75 in. X 22.75 in.)
 

3577A Product Page

 

Used Equipment Catalog

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved