| Accuracy applies over a temperature range of 23°C ± 5°C (at 0° to 55°C, error doubles). | |
| Parameters Measured | L-D, C-D R |
| Display | 3-1/2 digits max. display 1900 |
| Basic Accuracy | 0.2% |
| Test Frequency | 120 (100) Hz, 1 kHz ± 3% |
| Test Signal Level (Typical) | 1 V, 50 mV (Cp mode only) |
| DC Bias | Int. 1.5 V, 2.2 V, 6 V ± 5%,
selectable Ext. 0 to +30 V |
| Equivalent Circuit Modes | auto, parallel, series |
| Ranging Modes | LCR: auto, manual DQ: D only-fixed |
| Trigger | internal, external, manual |
| Measuring Terminal | 5-terminal configuration |
| Measuring Time (typical) | for a 1000 count measurement on a low loss component on a
fixed range: 1 kHz, 10 kHz: C/L 220-260 ms, R 120-160 ms 120 (100) Hz: C/L 900 ms, R 700 ms |
| Ranging Time | 1 kHz, 10 kHz: 180 ms/range step 120 (100) Hz: 670 ms/range step |
| Reading Rate | INT (internal trigger) approximately 30 ms between end of measurement cycle is initiated by a remote trigger input. |
| Accessories Available | 16061A: test fixture, direct couple, 5-terminal 16062A: test leads with alligator clips, 4-terminal (for low impedance measurements) 16063A: test leads with alligator clips, 3-terminal (for high impedance measurements) |
$20.00 Minimum on Total Order