<< UniSource Multimeters
UniSource B 4100
5½ Digits True RMS Programmable Bench Multimeter
|
- Full auto range, 5½ digits display (200,000
Counts)
- TRUE RMS AC Voltage Measurement
- Programmable Record (MAX / MIN)
- Programmable Relative (REF / ERR)
- Programmable Compare (GO-NOGO)
- Trend Plot : REF is the centerline value on
the trend graph.
- H/Div is the X axis (time) and V/Div is the Y
axis (amplitude).
- Setup : Setting and functions are selected
and stored in memory from the front panel. Eight
setups are stored in memory.
- Equipment : Display information about your
instrument including model number, serial number,
calibration date, GPIB address and firmware
revision.
- Remote Control : Built-in RS232C or optional
GPIB interface
- 128 by 64 pixel super twisted LCD display
- Power Source : 85V to 270VAC (±10%, 48 to
66Hz)
|
 
UniSource B4100 Shown
|
|
|
|
DC Voltage |
Range |
Resolution |
Accuracy |
Impedance |
Max. Input Voltage |
| 2.00000 V |
0.00001 V |
±(0.05%+3D) |
10 MΩ |
±1,000
V |
| 20.0000 V |
0.0001 V |
| 200.000 V |
0.001 V |
| 1000.00 V |
0.01 V |
0.1%±5D |
|
AC Voltage |
Range |
Resolution |
Accuracy |
Impedance |
Max. Input Voltage |
| 2.00000 V |
0.00001 V |
±(0.05%+3D) |
10 MΩ |
±1,000
V |
| 20.0000 V |
0.0001 V |
| 200.000 V |
0.001 V |
| 1000.00 V |
0.01 V |
0.1%±5D |
|
OHM |
Range |
Resolution |
Accuracy |
Max. Input Voltage |
| 200.000 Ω |
0.001 Ω |
±(0.2%+5D) |
±250 V |
| 2.00000 kΩ |
0.00001 kΩ |
±(0.05%+5D) |
| 20.0000 kΩ |
0.0001 kΩ |
| 200.000 kΩ |
0.001 kΩ |
| 2.00000 MΩ |
0.00001 MΩ |
| 10.0000 MΩ |
0.0001 MΩ |
±(0.2%+5D) |
Continuity
Buzzer |
100Ω or less |
|
Diode |
Range |
Resolution |
Max Test Current |
Max. Input Voltage |
| 2.00000 V |
0.00001 V |
About 1mA |
±250 V |
|
Frequency |
Range |
Resolution |
Accuracy |
Max. Input Voltage |
| 200.000 Hz |
0.001 Hz |
±(0.05%+2D) |
±250 V |
| 2.00000 kHz |
0.00001 kHz |
| 20.0000 kHz |
0.0001 kHz |
| 200.000 kHz |
0.001 kHz |
| 2.00000 MHz |
0.00001 MHz |
|
DC μA/mA |
Range |
Resolution |
Accuracy |
Max Burden |
Overload Protection |
| 200.000 μA |
0.001μA |
±(0.1%+5D) |
200 mV |
0.5A/250V Fuse |
| 200.000 mA |
0.001mA |
DC Ampere
(Max. 10A) |
Range |
Resolution |
Accuracy |
Max. Burden |
Overload Protection |
| 10.0000 A |
0.0001A |
±(0.1%+5D) |
200 mV |
0.5A/250V Fuse |
|
AC μA/mA |
Range |
Resolution |
Accuracy |
Max. Burden |
Overload Protection |
| 200.000μA |
0.01μA |
±(0.1%+5D) |
200 mV |
0.5A/250V Fuse |
| 200.000 mA |
0.01mA |
|
AC Ampere |
Range |
Resolution |
Accuracy |
Max. Burden |
Overload Protection |
| 10.0000 A |
0.001A |
±(0.1%+5D) |
200 mV |
10A /250V Fuse |
|
General |
Power Source |
85V to
270Vac(±10%, 48 to 66Hz) |
|
Dimensions / Weight |
235(W)
x 296(D) x 85(H) mm / About 1.5 kg |
|
Standard Accessories |
Users
Manual, Test lead, Line Cord,
Spare fuse, RS232 Cable and S/W |
|
Optional Accessories |
GP-IB
(IEEE-488.2) Interface (Installed) |
|
|
|
|