Specifications |
|
DC Voltage |
Ranges |
1000 mV, 10 V, 100 V, 1000 V |
|
Accuracy |
all ranges: ±(0.25 % rdg + 5 dgt) |
|
Resolution |
100 µV |
AC Voltage True RMS
(45 Hz To 2 kHz) |
Ranges |
1000 mV, 10 V, 100 V, 750 V |
|
Accuracy |
750 V range: 45 to 1 kHz: ±(2.0 % rdg + 10 dgts)
All other ranges, 45 to 500 Hz: ±(1.2 % rdg + 10 dgts)
All other ranges, 500 to 2 kHz: ±(2.0 % rdg + 10 dgts) |
AC + DC Voltage
(45 Hz To 2 kHz) |
Ranges |
1000 mV, 10 V, 100 V, 750 V |
|
Accuracy |
750 V range: 45 to 1 kHz: ±(2.5 % rdg + 10 dgts)
All other ranges, 45 to 500 Hz: ±(1.5 % rdg + 10 dgts)
All other ranges, 500 to 2 kHz: ±(2.5 % rdg + 10 dgts) |
|
Resolution |
100 µV |
|
DC Current |
Ranges |
100 µA, 1000 uA, 10 mA, 100 mA, 400 mA, 10 A |
|
Accuracy |
100 µA range: ±(0.5 % rdg + 10 dgt)
1000 uA to 400 mA ranges: ±(0.5 % rdg + 5 dgts)
10 A range: ±(1.5 % rdg + 10 dgt) |
|
Resolution |
0.01 µA |
AC Current (True RMS)
(AC + DC) (45 Hz - 1kHz) |
Ranges |
100 µA, 1000 µA, 10 mA, 100 mA, 400 mA, 10 A |
|
Accuracy |
100 µA to 100 mA ranges: ±(1.5 % rdg + 20 dgts)
400 mA range: ±(2.0 % rdg + 10 dgts)
10 A range: ±(2.5 % rdg + 20 dgts) |
|
Resolution |
0.01 µA |
|
AC coupled True RMS specified from 5 % to 100 % of range |
|
AC + DC Accuracy |
100 µA to 400 mA ranges: ±(2.0 % rdg + 20 dgts)
10 A range: ±(3.0 % rdg + 20 dgts) |
|
AC + DC: AC coupled True RMS specified from 10 % to 100 % of range |
|
Resistance |
Ranges |
1000Ω, 10kΩ, 100kΩ, 1000kΩ, 10MΩ, 40MΩ |
|
Accuracy |
1000Ω to 1000kΩ ranges: ±(0.5 % rdg + 8 dgts)
10MΩ range: ±(1.0 % rdg + 10 dgts)
40MΩ range: ±(2.0 % rdg + 10 dgts) |
|
Resolution |
100mΩ |
|
Capacitance |
Ranges |
40 nF, 400 nF, 4 µF, 40 µF, 400 µF |
|
Accuracy |
40 nF, 400 µF ranges: ±(3.0 % rdg + 10 dgts)
400 nF to 40 µF ranges: ±(3.0 % rdg + 5 dgts) |
|
Resolution |
0.01 nF |
|
Temperature |
Ranges |
-20 °C to 1300 °C, -4 °F to 2372 °F |
|
Accuracy |
-20 °C to 10 °C: ±(2.0 % rdg + 4 °C ) |
|
10 °C to 200 °C: ±(1.0 % rdg + 3 °C ) |
|
200 °C to 1300 °C: ±(2.0 % rdg + 2 °C ) |
|
-4 °F to 50 °F: ±(2.0 % rdg + 8 °F) |
|
50 °F to 400 °F: ±(1.0 % rdg + 6 °F) |
|
400 °F to 2372 °F: ±(2.0 % rdg + 4 °F) |
|
Resolution |
1 °C, 1 °F |
|
Frequency |
Ranges |
100 Hz, 1000 Hz, 10 kHz, 100 kHz, 1000 kHz, 10 MHz |
|
Accuracy |
all ranges: ±(0.1% rdg + 5 dgts) |
|
Resolution |
0.01 Hz |
|
Duty Cycle |
Ranges |
0 to 90 % |
|
Accuracy |
5 V logic: ±(2.0% rdg + 10 dgts) |
|
Resolution |
0.01 % |
|
Frequency Range |
0 % to 10 %: 40 Hz to 990 Hz
10 % to 90 %: 40 Hz to 20 kHz |
|
4-20mA DC Current Test |
Range |
0 – 100 % |
|
Accuracy |
±(0.5 % + 5 dgts) |
|
Resolution |
0.01 % |
|
dBm |
Ranges |
-13 dBm to + 50 dBm |
|
Accuracy |
45 Hz to 5 kHz: ±(0.7 dB + 8 dgts)
5 kHz to 10 kHz: ±(2.5 dB + 8 dgts) |
|
Resolution |
0.01 dBm |
|
Continuity |
Audible Indication |
< 40 Ω |
|
Diode Test |
Test Current |
0.5 mA (approximate) |
|
Open Circuit Volts |
3.0 V dc typical |
|
Peak |
Ac Volts Accuracy |
100 V, 750 V ranges: ±(3.0 + 200 dgts)
1000 mV, 10 V ranges: unspecified |
|
AC Current |
±(3.0 + 200 dgts) |
|
Overload Protection |
AC/DC Voltage |
µA/mA input: F0.5 A /1000 V fast-blow
ceramic 6.3 x 32 mm |
|
AC/DC Current |
10A input: F10 A /1000V fast-blow ceramic
10 x 38 mm |